Electrodynamometer Type Wattmeter

Encyclopedia
Field: Encyclopedia
0
China

Electrodynamometer Wattmeter Definition


An electrodynamometer type wattmeter measures electrical power by using the interaction between magnetic fields and electric currents.


Working Principle


Now let us look at constructional details of electrodynamometer. It consists of following parts.There are two types of coils present in the electrodynamometer. They are :


Moving Coil


The moving coil moves the pointer with the help of a spring control instrument. To prevent overheating, a limited current flows through the moving coil by connecting a high-value resistor in series. The air-cored moving coil is mounted on a pivoted spindle and can move freely. In an electrodynamometer type wattmeter, the moving coil acts as a pressure coil and is connected across the voltage, so the current through it is proportional to the voltage.


Fixed Coil


The fixed coil is divided into two equal parts and these are connected in series with the load, therefore the load current will flow through these coils. Now the reason is very obvious of using two fixed coils instead of one, so that it can be constructed to carry considerable amount of electric current. 


These coils are called the current coils of electrodynamometer type wattmeter. Earlier these fixed coils are designed to carry the current of about 100 amperes but now the modern wattmeter are designed to carry current of about 20 amperes in order to save power.


Control System


Out of two controlling systems i.e.


Gravity control


Spring control, only spring controlled systems are used in these types of wattmeter. Gravity controlled system cannot be employed because there will be appreciable amount of errors.


Damping System


Air friction damping is used because eddy current damping can distort the weak operating magnetic field, leading to errors.


There is uniform scale which is used in these types of instrument as moving coil moves linearly over a range of 40 degrees to 50 degrees on either side.


Now let us derive the expressions for the controlling torque and deflecting torques. In order to derive these expressions let us consider the circuit diagram given below:

 

9131cdae17853d6dfe3cfb3f249a7055.jpeg

 

We know that instantaneous torque in electrodynamic type instruments is directly proportional to the product of instantaneous values of currents flowing through both the coils and the rate of change of flux linked with the circuit.


Let I1 and I2 be the instantaneous values of currents in pressure and current coils respectively. So the expression for the torque can be written as:

 

65c01cfae06ca9a3843d154d4264ea11.jpeg

 

Where, x is the angle.


Now let the applied value of voltage across the pressure coil be

 

0749259f9178f078cbf0b88040d2f883.jpeg

 

Since the electrical resistance of the pressure coil is very high, we can neglect its reactance compared to its resistance. Thus, the impedance is equal to its electrical resistance, making it purely resistive.

The expression for instantaneous current can be written as I2 = v / Rp where Rp is the resistance of pressure coil.

 

6b25e9e95a562a7fff38ec3db617b544.jpeg

 

If there is phase difference between voltage and electric current, then expression for instantaneous current through current coil can be written as


2b23bb0b6bdb3bf7880bf08b00613dac.jpeg


As current through the pressure coil is very very small compared to the current through current coil hence current through the current coil can be considered as equal to total load current.Hence the instantaneous value of torque can be written as


851dc04955ce3a477f5f1ce1347a52d1.jpeg

Average value of deflecting torque can be obtained by integrating the instantaneous torque from limit 0 to T, where T is the time period of the cycle.


f51c520ea3dcf220e5eb60f3ac67d989.jpeg


Controlling torque is given by Tc = Kx where K is spring constant and x is final steady state value of deflection.


Advantages


  • Scale is uniform upto a certain limit.

  • They can be used for both to measure ac as well dc quantities as scale is calibrated for both.


Errors


  • Errors in the pressure coil inductance.



  • Errors may be due to pressure coil capacitance.



  • Errors may be due to mutual inductance effects.



  • Errors may be due connections.(i.e. pressure coil is connected after current coil)



  • Error due to Eddy currents.



  • Errors caused by vibration of moving system.



  • Temperature error.


  • Errors due to stray magnetic field.

Give a tip and encourage the author!

Recommended

Classification of Equipment Defects for Relay Protection and Safety Automatic Devices in Substations
In daily operations, various equipment defects are inevitably encountered. Whether maintenance personnel, operation and maintenance staff, or specialized management personnel, all must understand the defect classification system and adopt appropriate measures according to different situations.According to Q/GDW 11024-2013 "Operation and Management Guide for Relay Protection and Safety Automatic Devices in Smart Substations," equipment defects are classified into three levels based on severity an
12/15/2025
Under What Conditions Will the Line Circuit Breaker Auto-Reclosing Signal Be Locked Out?
The line circuit breaker auto-reclosing signal will be locked out if any of the following conditions occur:(1) Low SF6 gas pressure in circuit breaker chamber at 0.5MPa(2) Insufficient energy storage in circuit breaker operating mechanism or low oil pressure at 30MPa(3) Busbar protection operation(4) Circuit breaker failure protection operation(5) Line distance protection zone II or zone III operation(6) Short lead protection operation of circuit breaker(7) Presence of remote tripping signal(8)
12/15/2025
Application of Auto-Reclosing Residual Current Protective Devices in Lightning Protection for Communication Power Supplies
1. Power Interruption Problems Caused by RCD False Tripping During Lightning StrikesA typical communication power supply circuit is shown in Figure 1. A residual current device (RCD) is installed at the power supply input terminal. The RCD primarily provides protection against electrical equipment leakage currents to ensure personal safety, while surge protective devices (SPDs) are installed on power supply branches to protect against lightning intrusions. When lightning strikes occur, the senso
12/15/2025
Reclosing Charge Time: Why Does Reclosing Require Charging? What Effects Does Charging Time Have?
1. Function and Significance of Reclosing ChargingReclosing is a protective measure in power systems. After faults such as short circuits or circuit overloads occur, the system isolates the faulty circuit and then restores normal operation through reclosing. The function of reclosing is to ensure continuous operation of the power system, improving its reliability and safety.Before performing reclosing, the circuit breaker must be charged. For high-voltage circuit breakers, the charging time is g
12/15/2025
Send inquiry
+86
Click to upload file
Download
Get the IEE Business Application
Use the IEE-Business app to find equipment, obtain solutions, connect with experts, and participate in industry collaboration anytime, anywhere—fully supporting the development of your power projects and business.