What is Differential Voltmeter ?

Encyclopedia
Field: Encyclopedia
0
China

What is Differential Voltmeter ?

Definition: A voltmeter that measures the difference between a known voltage source and an unknown voltage source is called a differential voltmeter. It operates based on the principle of comparing the reference voltage source with the unknown voltage source.

The differential voltmeter offers very high accuracy. Its operating principle is similar to that of a potentiometer, and thus it is also known as a potentiometer voltmeter.

Construction of Differential Voltmeter
The circuit diagram of the differential voltmeter is shown below. A null meter is placed between the unknown voltage source and the precision divider. The output of the precision divider is connected to the known voltage source. The precision divider is adjusted until the null meter shows zero deflection.

Differential Voltmeter.jpg


When the meter shows zero deflection, it indicates that the magnitudes of both the known and the unknown voltage sources are equal. At this time of null deflection, neither the known nor the unknown source sends current to the meter, and the voltmeter presents high impedance to the measured source.

The null meter merely indicates the residual difference between the known and the unknown voltage sources. To accurately determine the difference between the sources, a more sensitive meter is employed.

A low - voltage DC standard source or a low - voltage Zener - controlled precision supply is utilized as the reference voltage. High - voltage supplies are used for measuring high voltages.

Types of Differential Voltmeter
There are two types of differential voltmeters:

  • AC Differential Voltmeter

  • DC Differential Voltmeter

AC Differential Voltmeter
The AC differential voltmeter is an enhanced version of DC instruments. The unknown AC voltage source is applied to a rectifier, which converts the AC voltage into a DC voltage of equivalent magnitude. The resulting DC voltage is then fed to a potentiometer for comparison with the known voltage source. The block diagram of the AC differential voltmeter is shown below.

Differential Voltmeter.jpg

The rectified AC voltage is compared with the standard DC voltage. When their magnitudes are equal, the meter shows null deflection. In this way, the value of the unknown voltage is determined.

DC Differential Voltmeter
The unknown DC source serves as the input to the amplifier section. A fraction of the output voltage is fed back to the input voltage by means of a divider network. Another part of the voltage divider provides a fractional input to the meter amplifier.

Differential Voltmeter.jpg

The meter is designed to measure the difference between the feedback voltage and the reference voltage. When the magnitudes of both the unknown voltage and the reference voltage are zero, the null meter indicates null deflection.

Give a tip and encourage the author!

Recommended

Classification of Equipment Defects for Relay Protection and Safety Automatic Devices in Substations
In daily operations, various equipment defects are inevitably encountered. Whether maintenance personnel, operation and maintenance staff, or specialized management personnel, all must understand the defect classification system and adopt appropriate measures according to different situations.According to Q/GDW 11024-2013 "Operation and Management Guide for Relay Protection and Safety Automatic Devices in Smart Substations," equipment defects are classified into three levels based on severity an
12/15/2025
Under What Conditions Will the Line Circuit Breaker Auto-Reclosing Signal Be Locked Out?
The line circuit breaker auto-reclosing signal will be locked out if any of the following conditions occur:(1) Low SF6 gas pressure in circuit breaker chamber at 0.5MPa(2) Insufficient energy storage in circuit breaker operating mechanism or low oil pressure at 30MPa(3) Busbar protection operation(4) Circuit breaker failure protection operation(5) Line distance protection zone II or zone III operation(6) Short lead protection operation of circuit breaker(7) Presence of remote tripping signal(8)
12/15/2025
Application of Auto-Reclosing Residual Current Protective Devices in Lightning Protection for Communication Power Supplies
1. Power Interruption Problems Caused by RCD False Tripping During Lightning StrikesA typical communication power supply circuit is shown in Figure 1. A residual current device (RCD) is installed at the power supply input terminal. The RCD primarily provides protection against electrical equipment leakage currents to ensure personal safety, while surge protective devices (SPDs) are installed on power supply branches to protect against lightning intrusions. When lightning strikes occur, the senso
12/15/2025
Reclosing Charge Time: Why Does Reclosing Require Charging? What Effects Does Charging Time Have?
1. Function and Significance of Reclosing ChargingReclosing is a protective measure in power systems. After faults such as short circuits or circuit overloads occur, the system isolates the faulty circuit and then restores normal operation through reclosing. The function of reclosing is to ensure continuous operation of the power system, improving its reliability and safety.Before performing reclosing, the circuit breaker must be charged. For high-voltage circuit breakers, the charging time is g
12/15/2025
Send inquiry
+86
Click to upload file
Download
Get the IEE Business Application
Use the IEE-Business app to find equipment, obtain solutions, connect with experts, and participate in industry collaboration anytime, anywhere—fully supporting the development of your power projects and business.