Frequency Limitation of an Oscilloscope

Encyclopedia
Field: Encyclopedia
0
China

Bandwidth Limit


Oscilloscopes, like multimeter, are essential tools for understanding circuits. However, they have limitations. To use an oscilloscope effectively, it’s crucial to know these limitations and find ways to address them.


A key feature of an oscilloscope is its bandwidth. Bandwidth determines how fast it can sample analog signals. What is bandwidth? Many think it’s the maximum frequency a scope can handle. Actually, bandwidth is the frequency where a signal’s amplitude drops by 3dB, or 29.3% below its true amplitude.


At the maximum rated frequency, the oscilloscope shows 70.7% of the signal’s actual amplitude. For example, if the actual amplitude is 5V, the scope will display it as about 3.5V.

 

1db1cd3ca65bcd1fd6337fbcfd0c9fdd.jpeg

 

Oscilloscopes with a bandwidth of 1 GHz or less have a Gaussian or low-pass frequency response, starting at one-third of the -3 dB frequency and gradually decreasing at higher frequencies.


Scopes with specification greater than 1 GHz shows a maximally flat response with a sharper roll-off near the -3dB frequency. The lowest frequency of oscilloscope at which the input signal is attenuated by 3 dB is considered as the bandwidth of the scope. The oscilloscope with a maximally flat response can attenuate in-band signals which are less as comparative to the oscilloscope with the Gaussian response and do more accurate measurements on in-band signals.


On the other hand, the scope with Gaussian response attenuated out-bands signals which are less comparative to the scope with the maximally flat response. It means that such scope has a faster rise time comparative to other scopes with same bandwidth specification. Rise time specification of a scope is closely related to its bandwidth.


A Gaussian response type oscilloscope will have a rise time of 0.35/f BW approximately based on a 10% to 90% criterion. A maximally flat response type scope has a rise time of 0.4/f BW approximately based on the sharpness of the frequency roll-off characteristic.

 

Rise time is the fastest edge speed an oscilloscope can display if the input signal has an infinitely fast rise time. Measuring this theoretical value is impossible, so it’s better to calculate a practical value

 

e425a56d91e632e215aecb99858dbbe3.jpeg

 

Precaution Required for Precise Measurements in Oscilloscope


The foremost thing that users must know is the bandwidth limitation of the scope. The bandwidth of the oscilloscope should be wide enough to accommodate the frequencies within the signal and display the waveform properly.


The probe used with the scope played an important role in the performance of the equipment. The bandwidth of the oscilloscope, as well as probe, should be in proper combination. Using an improper oscilloscope probe can spoil the performance of the entire test equipment.


To measure frequency as well as amplitude accurately, the bandwidth of both the scope and the probe attached to it are well above the signal you want to capture precisely. For example, if the required accuracy of amplitude is to ~1%, then berate factor of scope by 0.1x, that means 100MHz scope can capture 10MHz with a 1% error in amplitude.


One must take into consideration the correct triggering of the scope so that the resultant view of the waveform is much clearer.


Users should be aware of ground clips while taking high-speed measurements. The wire of the clip produces inductance and ringing into the circuit which affects the measurements.


The summary of the whole article is that for analog scope, the bandwidth of the scope is at least, three times higher than the highest analog frequency of the system. For the digital application, the bandwidth of the scope is at least, five times higher than the fastest clock rate of the system. 


Give a tip and encourage the author!

Recommended

Classification of Equipment Defects for Relay Protection and Safety Automatic Devices in Substations
In daily operations, various equipment defects are inevitably encountered. Whether maintenance personnel, operation and maintenance staff, or specialized management personnel, all must understand the defect classification system and adopt appropriate measures according to different situations.According to Q/GDW 11024-2013 "Operation and Management Guide for Relay Protection and Safety Automatic Devices in Smart Substations," equipment defects are classified into three levels based on severity an
12/15/2025
Under What Conditions Will the Line Circuit Breaker Auto-Reclosing Signal Be Locked Out?
The line circuit breaker auto-reclosing signal will be locked out if any of the following conditions occur:(1) Low SF6 gas pressure in circuit breaker chamber at 0.5MPa(2) Insufficient energy storage in circuit breaker operating mechanism or low oil pressure at 30MPa(3) Busbar protection operation(4) Circuit breaker failure protection operation(5) Line distance protection zone II or zone III operation(6) Short lead protection operation of circuit breaker(7) Presence of remote tripping signal(8)
12/15/2025
Application of Auto-Reclosing Residual Current Protective Devices in Lightning Protection for Communication Power Supplies
1. Power Interruption Problems Caused by RCD False Tripping During Lightning StrikesA typical communication power supply circuit is shown in Figure 1. A residual current device (RCD) is installed at the power supply input terminal. The RCD primarily provides protection against electrical equipment leakage currents to ensure personal safety, while surge protective devices (SPDs) are installed on power supply branches to protect against lightning intrusions. When lightning strikes occur, the senso
12/15/2025
Reclosing Charge Time: Why Does Reclosing Require Charging? What Effects Does Charging Time Have?
1. Function and Significance of Reclosing ChargingReclosing is a protective measure in power systems. After faults such as short circuits or circuit overloads occur, the system isolates the faulty circuit and then restores normal operation through reclosing. The function of reclosing is to ensure continuous operation of the power system, improving its reliability and safety.Before performing reclosing, the circuit breaker must be charged. For high-voltage circuit breakers, the charging time is g
12/15/2025
Send inquiry
+86
Click to upload file
Download
Get the IEE Business Application
Use the IEE-Business app to find equipment, obtain solutions, connect with experts, and participate in industry collaboration anytime, anywhere—fully supporting the development of your power projects and business.